Skip to main content
Design & Integration

Options for Exhausting Air From Units

The three exhaust scenario options are described below:

1. Exhausting out a standard window with the provided accessories.

a. Open the window and place the vent assembly into place with external cowling facing downward for weather protection. The wings of the vent can be extended to fit the length of the window opening. The exhaust hose is connected from the unit to the vent assembly with an included clamp.

i. Dimensions of the vent are Height: 10 inches, Total Depth: 5 – 3/8 inches, Panel Depth (part held in place by window): 1 – 1/8 inches, Length (adjustable): 24 to 36 inches

2. The vent assembly can also be used in place of a commercial ceiling tile, exhausting treated air into the space above the ceiling. The telescoping vent assembly will fit into many ceiling tile set ups. Note: the remaining gap in the ceiling tile space is then filled by cutting the existing ceiling tile to the appropriate size and re-fitting into that gap. The exhaust hose is then connected to the vent assembly with the included clamp.

3. The exhaust hose can also be connected directly into the existing HVAC system exhaust. It is recommended to contact a local HVAC professional for custom HVAC installations.